Current Step

1"Level phantom on table. Align CAX to crosshair. Use mechanical front pointer to set phantom crosshair at 100 cm SSD. 測試 固定品保膜體在治療床上, 使用CAX與Front Pointer確認膜體的中心點在100cm SSD."
     

2"Rotate gantry to 90° and verify CAX falls between short horizontal marks (indicates <1° gantry angle readout variance). A2-12旋轉機架至90度, 並且確認CAX落在兩條短水平線間(表示誤差在1度以內)"