Skip Navigation Links

TG142 Demo

Site:  Machine: 
>1<
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
Current Step

1Level phantom on table. Align CAX to crosshair. Use mechanical front pointer to set phantom crosshair at 100 cm SSD.
     

2Rotate gantry to 90° and verify CAX falls between short horizontal marks (indicates <1° gantry angle readout variance).