Skip Navigation Links

Report from iQA Using CT-SIM/Check


Site: Demo Site
Machine: CT SIM
--
User Data Entry Completed:2022-05-04 21:18:38
Imaged: 2/15/2022 8:58 AM
Overall Status:Failed
1.FailedHU Calibration (Cylinder Mean)55.32Image for CylinderMean  Trend
2.PassedHU Uniformity (Cylinder Standard Deviation)11.21Image for CylinderStdDev  Trend
3.PassedImage OrientationHFSImage for ScanOrient  Image for ScanOrient  Trend
4.PassedPlug 1 HU Mean-686.17Image for P1Mean  Trend
5.PassedPlug 2 HU Mean-488.10Image for P2Mean  Trend
6.PassedPlug 3 HU Mean668.87Image for P3Mean  Trend
7.PassedPlug 4 HU Mean60.42Image for P4Mean  Trend
8.PassedPlug 4 Inner HU Mean58.24Image for P4InnerMean  Trend
9.PassedPlug 4 Inner StDev9.12Image for P4InnerStdDev  Trend
10.PassedScan Center Lateral (mm)0.1Image for CtrX  Image for CtrX  Image for CtrX  Trend
11.PassedScan Center Vertical (mm)-128.5Image for CtrY  Image for CtrY  Trend
12.PassedScan Center Longitudinal (mm)-594.8Image for CtrZ  Image for CtrZ  Trend
13.PassedLateral Scale (%)100.0Image for ScaleX  Trend
14.PassedVertical Scale (%)99.3Image for ScaleY  Trend
15.PassedLongitudinal Scale (% Ratio of Slice Spacing)100.0Image for ScaleZ  Trend
16.PassedWire Width Ratio1.592Image for WireWidthRatio  Trend
17.PassedPitch (°)0.0Image for RotX  Trend
18.PassedYaw (°)-0.1Image for RotY  Trend
19.PassedRoll (°)0.1Image for RotZ  Trend
20.PassedTarget Pt Left Tower, Lateral (mm)-178.4Image for TgtOffX  Trend
21.PassedTarget Pt Left Tower, Vertical (mm)149.5Image for TgtOffY  Trend
22.PassedTarget Pt Left Tower, Longitudinal (mm)49.2Image for TgtOffZ  Trend

*** End of Protocol ***