Skip Navigation Links

Report from iQA Using CT-SIM/Check


Site: Demo Site
Machine: CT SIM
--
User Data Entry Completed:2022-01-21 23:37:13
Imaged: 10/7/2021 12:33 PM - 12:34 PM
Overall Status:Marginal
1.PassedHU Calibration (Cylinder Mean)74.31Image for CylinderMean  Trend
2.PassedHU Uniformity (Cylinder Standard Deviation)4.75Image for CylinderStdDev  Trend
3.PassedImage OrientationHFSImage for ScanOrient  Image for ScanOrient  Trend
4.PassedPlug 1 HU Mean680.38Image for P1Mean  Trend
5.PassedPlug 2 HU Mean-504.35Image for P2Mean  Trend
6.PassedPlug 3 HU Mean-706.60Image for P3Mean  Trend
7.PassedPlug 4 HU Mean78.63Image for P4Mean  Trend
8.PassedPlug 4 Inner HU Mean78.19Image for P4InnerMean  Trend
9.PassedPlug 4 Inner StDev4.00Image for P4InnerStdDev  Trend
10.MarginalScan Center Lateral (mm)-1.4Image for CtrX  Image for CtrX  Image for CtrX  Trend
11.PassedScan Center Vertical (mm)-0.3Image for CtrY  Image for CtrY  Trend
12.PassedScan Center Longitudinal (mm)0.6Image for CtrZ  Image for CtrZ  Trend
13.PassedLateral Scale (%)100.1Image for ScaleX  Trend
14.PassedVertical Scale (%)100.1Image for ScaleY  Trend
15.MarginalLongitudinal Scale (% Ratio of Slice Spacing)94.6Image for ScaleZ  Trend
16.PassedWire Width Ratio1.111Image for WireWidthRatio  Trend
17.PassedPitch (°)-0.1Image for RotX  Trend
18.PassedYaw (°)0.1Image for RotY  Trend
19.PassedRoll (°)0.1Image for RotZ  Trend
20.PassedTarget Pt Left Tower, Lateral (mm)-178.1Image for TgtOffX  Trend
21.PassedTarget Pt Left Tower, Vertical (mm)149.8Image for TgtOffY  Trend
22.PassedTarget Pt Left Tower, Longitudinal (mm)49.2Image for TgtOffZ  Trend

*** End of Protocol ***