Skip Navigation Links

Report from iQA Using CT-SIM/Check


Site: Demo Site
Machine: CT SIM
--
User Data Entry Completed:2021-11-03 15:34:53
Imaged: 10/26/2021 10:03 AM
Overall Status:Failed
1.PassedHU Calibration (Cylinder Mean)75.11Image for CylinderMean  Trend
2.PassedHU Uniformity (Cylinder Standard Deviation)13.20Image for CylinderStdDev  Trend
3.PassedImage OrientationHFSImage for ScanOrient  Image for ScanOrient  Trend
4.PassedPlug 1 HU Mean-696.01Image for P1Mean  Trend
5.PassedPlug 2 HU Mean-494.31Image for P2Mean  Trend
6.PassedPlug 3 HU Mean59.15Image for P3Mean  Trend
7.PassedPlug 4 HU Mean660.68Image for P4Mean  Trend
8.PassedPlug 4 Inner HU Mean649.62Image for P4InnerMean  Trend
9.PassedPlug 4 Inner StDev9.71Image for P4InnerStdDev  Trend
10.PassedScan Center Lateral (mm)0.0Image for CtrX  Image for CtrX  Image for CtrX  Trend
11.PassedScan Center Vertical (mm)-115.4Image for CtrY  Image for CtrY  Trend
12.PassedScan Center Longitudinal (mm)-0.4Image for CtrZ  Image for CtrZ  Trend
13.PassedLateral Scale (%)100.2Image for ScaleX  Trend
14.PassedVertical Scale (%)100.1Image for ScaleY  Trend
15.PassedLongitudinal Scale (% Ratio of Slice Spacing)99.7Image for ScaleZ  Trend
16.PassedWire Width Ratio - Sensitivity Profile1.286Image for WireWidthRatio  Trend
17.PassedPitch (°)0.2Image for RotX  Trend
18.PassedYaw (°)0.0Image for RotY  Trend
19.PassedRoll (°)0.2Image for RotZ  Trend
20.PassedTarget Pt Left Tower, Lateral (mm)-178.3Image for TgtOffX  Trend
21.PassedTarget Pt Left Tower, Vertical (mm)149.2Image for TgtOffY  Trend
22.FailedTarget Pt Left Tower, Longitudinal (mm)50.5Image for TgtOffZ  Trend

*** End of Protocol ***