Skip Navigation Links

Report from iQA Using CT-SIM/Check


Site: Demo Site
Machine: CT SIM
--
User Data Entry Completed:2021-11-02 16:46:30
Imaged: 10/18/2021 7:25 PM
Overall Status:Failed
1.PassedHU Calibration (Cylinder Mean)13.41Image for CylinderMean  Trend
2.PassedHU Uniformity (Cylinder Standard Deviation)9.15Image for CylinderStdDev  Trend
3.PassedImage OrientationHFSImage for ScanOrient  Image for ScanOrient  Trend
4.PassedPlug 1 HU Mean-698.63Image for P1Mean  Trend
5.PassedPlug 2 HU Mean-490.16Image for P2Mean  Trend
6.PassedPlug 3 HU Mean-2.09Image for P3Mean  Trend
7.PassedPlug 4 HU Mean686.31Image for P4Mean  Trend
8.PassedPlug 4 Inner HU Mean688.50Image for P4InnerMean  Trend
9.PassedPlug 4 Inner StDev9.03Image for P4InnerStdDev  Trend
10.PassedScan Center Lateral (mm)0.3Image for CtrX  Image for CtrX  Image for CtrX  Trend
11.PassedScan Center Vertical (mm)-0.2Image for CtrY  Image for CtrY  Trend
12.FailedScan Center Longitudinal (mm)-914.9Image for CtrZ  Image for CtrZ  Trend
13.PassedLateral Scale (%)100.2Image for ScaleX  Trend
14.PassedVertical Scale (%)99.6Image for ScaleY  Trend
15.FailedLongitudinal Scale (% Ratio of Slice Spacing)92.4Image for ScaleZ  Trend
16.PassedWire Width Ratio - Sensitivity Profile1.205Image for WireWidthRatio  Trend
17.PassedPitch (°)0.0Image for RotX  Trend
18.PassedYaw (°)0.0Image for RotY  Trend
19.PassedRoll (°)-0.2Image for RotZ  Trend
20.PassedTarget Pt Left Tower, Lateral (mm)-178.4Image for TgtOffX  Trend
21.PassedTarget Pt Left Tower, Vertical (mm)150.2Image for TgtOffY  Trend
22.FailedTarget Pt Left Tower, Longitudinal (mm)50.2Image for TgtOffZ  Trend

*** End of Protocol ***