Skip Navigation Links

Report from iQA Using CT-SIM/Check


Site: Demo Site
Machine: CT SIM
--
User Data Entry Completed:2021-10-22 02:34:03
Imaged: 10/19/2021 8:56 AM
Overall Status:Failed
1.PassedHU Calibration (Cylinder Mean)58.61Image for CylinderMean  Trend
2.PassedHU Uniformity (Cylinder Standard Deviation)10.05Image for CylinderStdDev  Trend
3.PassedImage OrientationHFSImage for ScanOrient  Image for ScanOrient  Trend
4.PassedPlug 1 HU Mean-719.94Image for P1Mean  Trend
5.PassedPlug 2 HU Mean-504.69Image for P2Mean  Trend
6.PassedPlug 3 HU Mean748.33Image for P3Mean  Trend
7.PassedPlug 4 HU Mean88.46Image for P4Mean  Trend
8.PassedPlug 4 Inner HU Mean80.79Image for P4InnerMean  Trend
9.PassedPlug 4 Inner StDev2.82Image for P4InnerStdDev  Trend
10.PassedScan Center Lateral (mm)-0.1Image for CtrX  Image for CtrX  Image for CtrX  Trend
11.PassedScan Center Vertical (mm)-128.5Image for CtrY  Image for CtrY  Trend
12.FailedScan Center Longitudinal (mm)-594.9Image for CtrZ  Image for CtrZ  Trend
13.PassedLateral Scale (%)100.0Image for ScaleX  Trend
14.PassedVertical Scale (%)99.8Image for ScaleY  Trend
15.PassedLongitudinal Scale (% Ratio of Slice Spacing)100.5Image for ScaleZ  Trend
16.PassedWire Width Ratio - Sensitivity Profile1.559Image for WireWidthRatio  Trend
17.PassedPitch (°)0.0Image for RotX  Trend
18.PassedYaw (°)0.0Image for RotY  Trend
19.PassedRoll (°)0.0Image for RotZ  Trend
20.PassedTarget Pt Left Tower, Lateral (mm)-177.8Image for TgtOffX  Trend
21.MarginalTarget Pt Left Tower, Vertical (mm)148.9Image for TgtOffY  Trend
22.FailedTarget Pt Left Tower, Longitudinal (mm)50.0Image for TgtOffZ  Trend

*** End of Protocol ***