Site: Demo Site
Machine: CT SIM
--
User Data Entry Completed:2021-06-03 23:59:38
Imaged: 5/19/2021 1:42 PM
Overall Status:Failed
1.FailedHU Calibration (Cylinder Mean)-990.73Image for CylinderMean  Trend
2.FailedHU Uniformity (Cylinder Standard Deviation)60.13Image for CylinderStdDev  Trend
3.PassedImage OrientationHFSImage for ScanOrient  Image for ScanOrient  Trend
4.PassedPlug 1 HU Mean-688.37Image for P1Mean  Trend
5.PassedPlug 2 HU Mean-490.71Image for P2Mean  Trend
6.PassedPlug 3 HU Mean672.25Image for P3Mean  Trend
7.PassedPlug 4 HU Mean62.62Image for P4Mean  Trend
8.PassedPlug 4 Inner HU Mean62.02Image for P4InnerMean  Trend
9.PassedPlug 4 Inner StDev16.22Image for P4InnerStdDev  Trend
10.PassedScan Center Lateral (mm)-0.4Image for CtrX  Image for CtrX  Image for CtrX  Trend
11.PassedScan Center Vertical (mm)-128.4Image for CtrY  Image for CtrY  Trend
12.FailedScan Center Longitudinal (mm)-594.2Image for CtrZ  Image for CtrZ  Trend
13.PassedLateral Scale (%)100.1Image for ScaleX  Trend
14.MarginalVertical Scale (%)98.5Image for ScaleY  Trend
15.FailedLongitudinal Scale (% Ratio of Slice Spacing)105.2Image for ScaleZ  Trend
16.PassedWire Width Ratio - Sensitivity Profile1.537Image for WireWidthRatio  Trend
17.PassedPitch (°)-0.1Image for RotX  Trend
18.PassedYaw (°)-0.2Image for RotY  Trend
19.PassedRoll (°)0.4Image for RotZ  Trend
20.PassedTarget Pt Left Tower, Lateral (mm)-178.8Image for TgtOffX  Trend
21.MarginalTarget Pt Left Tower, Vertical (mm)148.7Image for TgtOffY  Trend
22.PassedTarget Pt Left Tower, Longitudinal (mm)50.6Image for TgtOffZ  Trend

*** End of Protocol ***