Skip Navigation Links

Report from iQA Using CT-SIM/Check


Site: Demo Site
Machine: CT SIM
--
User Data Entry Completed:2021-06-01 23:17:28
Imaged: 6/2/2021 10:11 AM
Overall Status:Failed
1.PassedHU Calibration (Cylinder Mean)53.39Image for CylinderMean  Trend
2.PassedHU Uniformity (Cylinder Standard Deviation)8.88Image for CylinderStdDev  Trend
3.PassedImage OrientationHFSImage for ScanOrient  Image for ScanOrient  Trend
4.PassedPlug 1 HU Mean-689.01Image for P1Mean  Trend
5.PassedPlug 2 HU Mean-490.04Image for P2Mean  Trend
6.PassedPlug 3 HU Mean673.74Image for P3Mean  Trend
7.PassedPlug 4 HU Mean64.09Image for P4Mean  Trend
8.PassedPlug 4 Inner HU Mean62.08Image for P4InnerMean  Trend
9.PassedPlug 4 Inner StDev5.38Image for P4InnerStdDev  Trend
10.PassedScan Center Lateral (mm)-30.2Image for CtrX  Image for CtrX  Image for CtrX  Trend
11.PassedScan Center Vertical (mm)-128.5Image for CtrY  Image for CtrY  Trend
12.FailedScan Center Longitudinal (mm)-595.0Image for CtrZ  Image for CtrZ  Trend
13.PassedLateral Scale (%)100.0Image for ScaleX  Trend
14.PassedVertical Scale (%)100.3Image for ScaleY  Trend
15.PassedLongitudinal Scale (% Ratio of Slice Spacing)100.2Image for ScaleZ  Trend
16.PassedWire Width Ratio - Sensitivity Profile1.909Image for WireWidthRatio  Trend
17.PassedPitch (°)-0.2Image for RotX  Trend
18.PassedYaw (°)0.0Image for RotY  Trend
19.PassedRoll (°)0.0Image for RotZ  Trend
20.PassedTarget Pt Left Tower, Lateral (mm)-178.1Image for TgtOffX  Trend
21.PassedTarget Pt Left Tower, Vertical (mm)149.5Image for TgtOffY  Trend
22.MarginalTarget Pt Left Tower, Longitudinal (mm)48.7Image for TgtOffZ  Trend

*** End of Protocol ***