Skip Navigation Links

Report from iQA Using CT-SIM/Check


Site: Demo Site
Machine: CT SIM
--
User Data Entry Completed:2021-04-30 04:51:30
Imaged: 4/28/2021 9:09 AM
Overall Status:Failed
1.PassedHU Calibration (Cylinder Mean)56.29Image for CylinderMean  Trend
2.PassedHU Uniformity (Cylinder Standard Deviation)8.20Image for CylinderStdDev  Trend
3.PassedImage OrientationHFSImage for ScanOrient  Image for ScanOrient  Trend
4.PassedPlug 1 HU Mean-691.40Image for P1Mean  Trend
5.PassedPlug 2 HU Mean-491.39Image for P2Mean  Trend
6.PassedPlug 3 HU Mean675.46Image for P3Mean  Trend
7.PassedPlug 4 HU Mean62.61Image for P4Mean  Trend
8.PassedPlug 4 Inner HU Mean60.53Image for P4InnerMean  Trend
9.PassedPlug 4 Inner StDev4.79Image for P4InnerStdDev  Trend
10.PassedScan Center Lateral (mm)-0.5Image for CtrX  Image for CtrX  Image for CtrX  Trend
11.PassedScan Center Vertical (mm)-127.2Image for CtrY  Image for CtrY  Trend
12.FailedScan Center Longitudinal (mm)-594.7Image for CtrZ  Image for CtrZ  Trend
13.PassedLateral Scale (%)100.2Image for ScaleX  Trend
14.PassedVertical Scale (%)100.7Image for ScaleY  Trend
15.FailedLongitudinal Scale (% Ratio of Slice Spacing)92.6Image for ScaleZ  Trend
16.PassedWire Width Ratio - Sensitivity Profile1.568Image for WireWidthRatio  Trend
17.PassedPitch (°)-0.3Image for RotX  Trend
18.PassedYaw (°)0.0Image for RotY  Trend
19.PassedRoll (°)0.3Image for RotZ  Trend
20.PassedTarget Pt Left Tower, Lateral (mm)-178.1Image for TgtOffX  Trend
21.MarginalTarget Pt Left Tower, Vertical (mm)148.7Image for TgtOffY  Trend
22.MarginalTarget Pt Left Tower, Longitudinal (mm)48.9Image for TgtOffZ  Trend

*** End of Protocol ***